Figure 7: Capacitance and AC voltage
A8. As mentioned earlier, as the capacitance and frequency increase, the impedance of the DUT will decrease. To further illustrate this point, the following example shows by calculation how the impedance of the DUT affects the actual voltage seen on the DUT.
Use HP 4263B tester and HP4278A tester to test capacitors (C3216Y5V1A106Z). Putting a frequency of 1kHz and a capacitance of 10 µF into Equation 1 produces a capacitor impedance of approximately 16 Ω.
When 1.0 Vrms should be added to the capacitor from the test equipment, the voltage is divided between the meter impedance and the DUT impedance. The figure shows that the impedance of HP4263B remains 100 Ω, but for HP4278A, the calculated capacitor impedance becomes 1.5 Ω. As a result, for HP4263B, most of the applied voltage drops in the tester's impedance, while HP4278A enables the capacitor to receive most of the voltage (Figure 5-6). The result is that the indicated value displayed by HP4263B will be lower than the true value.
As shown above, setting the OSC voltage of the test instrument to 1.0Vrms does not guarantee that all external voltages will be printed on the DUT. It is not surprising that the voltage on the DUT is about 10% of the set value. Figure 7 shows how low VAC affects the measured capacitance.
The figure below shows the difference between ALC on and off. When the ALC function is off, the actual voltage on the DUT is about 10% of the set voltage of 1.0Vrms. When the ALC function is on, the voltage on the DUT is almost 100% of the set voltage. HP4284A has voltage and current level monitors, but you can still use a voltmeter to verify the actual voltage.
ALC function of HP4284A capacitance meter

